16.
The current passing through a resistance of 100 ± 0.2% Ω is 2 ± 0.5% A. The limiting error in computation of power will be
Answer: Option
Explanation:
Limiting error = 2 x 0.5 + 0.2 = ± 1.2% (we have to take 2 x 0.5 because power ∝ I2).
17.
If the secondary burden of a CT 15 VA and secondary current is 5 A, the load impedance of CT is
Answer: Option
Explanation:
.
18.
Consider the following instruments
- Tangent galvanometer
- Rayleigh current balance
- Household energy meter
- Hot wire ammeter
Answer: Option
Explanation:
Only tangent galvanometer and Rayleigh Current balance are absolute instruments. Energy meter and ammeter are secondary instruments.
19.
A 0-150 voltmeter has a guaranteed accuracy of 1% of full scale reading. If actual reading is 83 V, the percentage error is
Answer: Option
Explanation:
.
20.
In a CRO, the frequency of output of time base generator
Answer: Option
Explanation:
Time base signal frequency is adjustable so that voltages of different frequencies can be measured.