6.
The most commonly used moving iron instruments are
Answer: Option
Explanation:
Repulsion type have better accuracy.
7.
A moving iron instrument can be used for
Answer: Option
Explanation:
Its deflecting torque ∝ I2. So, it can be used for both d.c. and a.c.
8.
Two voltmeters are each 0-300 V range. Their internal resistances are 10 kΩ and 20 kΩ They are connected is series and a voltage of 400 V is applied. Their readings respectively will be
Answer: Option
Explanation:
The readings will be proportional to their internal resistance,
.
9.
A 300 V voltmeter has an accuracy of ± 2% of full scale deflection. When the reading is 222 V, the actual voltage
Answer: Option
Explanation:
Error = ± 6 V. Therefore, the actual value would be between 222 ± 6 V.
10.
In a thermocouple two metal junctions between metals M1 and M2 are kept at temperature T1 and T2. The thermocouple emf is produced because
Answer: Option
Explanation:
The two materials in thermocouple are different and the two junctions are kept at different temperatures.